  "dataset": "AeroSense plate P2, instrument runs 904–935",
  "acquisition_dates_from_headers": ["2026-10-10", "2026-10-11"],
  "confirmation_date": "2026-10-11",
  "confirmed_by_team_in_conversation": {
    "cell_system": "HEK293",
    "time_after_transfection": "approximately 72 hours",
    "G418_selection": "none before this measurement",
    "available_experimental_plates": 1,
    "VOC_final_concentrations": "unknown",
    "VUAA1_solvent": "DMSO",
    "final_DMSO_fraction": "unknown",
    "VOC_preparation": "initially aqueous; diluted into assay buffer for testing",
    "assay_buffer": {
      "NaCl_mM": 140,
      "KCl_mM": 5,
      "CaCl2_mM": 2,
      "MgCl2_mM": 1,
      "HEPES_Na_mM": 5,
      "glucose_mM": 20,
      "pH": 7.3,
      "osmolarity_mOsm": 340
    },
    "numeric_gain": "not known"
  },
  "source_report_only": {
    "VUAA1_nominal_final_concentration_uM": 50,
    "instrument": "BMG CLARIOstar Plus",
    "plate": "GREINER 96 F-BOTTOM",
    "optic": "bottom",
    "dichroic_filter": "LP 504",
    "focal_height_mm": -4.1,
    "flashes_per_well_per_cycle": 20,
    "settling_time_s": 0.5,
    "EDR": "off",
    "temperature_control": "off",
    "nominal_full_scale_RFU": 260000,
    "mixing": "no shaking during response recording",
    "addition": "manual 8-channel pipette; columns 1–8 then 9–12",
    "volume_sequence_uL": [100, 150, 200],
    "gain_mismatch_exclusions": "B both baseline–response pairs; C first pair",
    "cell_detachment": "reported for some P2 cells; not quantified"
  },
  "verified_in_raw_CSV": {
    "excitation": "482-16",
    "emission": "530-40",
    "baseline_readings_per_stage": 5,
    "response_readings_per_stage": 15,
    "cycle_interval_s": 12,
    "baseline_time_s": [0, 12, 24, 36, 48],
    "response_time_span_s": [0, 168]
  },
  "unresolved": [
    "Numeric gain for every run and confirmation of gain consistency",
    "VUAA1 final concentration independently verified from preparation records",
    "Final DMSO percentage and matched DMSO-only vehicle wells",
    "VOC doses, chemical identities/purity and preparation details",
    "Seeding density, passage, viability and quantified detachment",
    "Plasmid-batch identities and linkage to earlier microscopy cultures",
    "Exact addition-to-first-reading delays and environmental temperature"
  ],
  "replication": "One well per condition×stimulus on one available plate. Repeated timepoints are longitudinal readings, not biological or well replicates.",
  "replication_provenance": "One available plate was confirmed by the team. One well per condition×stimulus is derived from the supplied 8-stimulus by 12-condition layout and 96 distinct well identifiers in the instrument exports.",
  "interpretation": "Exploratory fluorescence observations; no membrane-localization, receptor-specificity, stable-line or validated detection-limit claim."
}